5
Troubleshooting
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
5-29
5.3.5 System shut
down messages
The system shut down message displays in the format
“System
shutdown
error
xxyy”
and is intended to be a help for complex failure analysis. The message consist of a
4 digit code
“ xxyy
”, that represents the last information status prior to the
microprocessor
systems
turn
down.
•
The “xx” represents the last “safe” status of micro controller A
•
The “yy” represents the last “safe” status of micro controller B
Error code Alarming uC Set condition
Possible failure
Action
Improper connection of
internal SIO cable
Check cable from front panel
bracket to X 14 on B board and
replace, if necessary
Improper connection of
external SIO cable (Control
panel cable) Or Connection
problem with external SIO
cable connector (Lemosa
plug)
Check front panel cable and
plug (Lemosa) and replace, if
necessary
Defective uC controller A or
Defective safety diode on PC
board A
Replace PC board A
11yy A
Timer
3
system
problem
Defective Fuse SI 21, SI 22
Check fuse SI 21 and SI 22
and replace, if necessary
13yy A
Timer
6
system
problem
Defective uC controller A
Replace PC board A
14yy
A
Front side routine
without function
Defective uC controller A
Replace PC board A
ESD or EMI problem
Check environment for ESD or
EMI source
Improper connection of
external SIO cable (Control
panel cable) Or Connection
problem with external SIO
cable connector (Lemosa
plug)
Check front panel cable and
plug (Lemosa) and replace, if
necessary
15yy
A
SIO data non plausible
Defective uC controller A
Replace PC board A
Содержание Centiva/5 Plus
Страница 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Страница 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Страница 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Страница 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Страница 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Страница 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Страница 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Страница 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Страница 178: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 2 9 1 Pneumatic diagram ...
Страница 180: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 4 9 2 Electrical diagram ...
Страница 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Страница 182: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 6 9 3 Component layout PC board B ...
Страница 183: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 7 9 4 Component layout PC board A ...
Страница 192: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 16 Notes ...
Страница 206: ...10 Configuration and optionalization TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 10 14 Notes ...
Страница 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Страница 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...