Error code Alarming uC Set condition
Possible failure
Action
xx10
B
AD converter defective Defective uC controller B
Replace PC board B
Improper connection of
internal SIO cable
Check cable from front panel
bracket to X 14 on B board and
replace, if necessary
Improper connection of
external SIO cable (Control
panel cable) Or Connection
problem with external SIO
cable connector (Lemosa
plug)
Check front panel cable and
plug (Lemosa) and replace, if
necessary
xx11 B
Timer
3
system
problem
Defective uC controller B
Replace PC board B
xx12 B
Timer
5
system
problem
Defective uC controller B
Replace PC board B
xx13 B
Timer
6
system
problem
Defective uC controller B
Replace PC board B
Defective uC controller B
Replace PC board B
xx14 B
Front
side
routine
without function
Software down
Restart and check
ESD or EMI problems
Check environment for ESD or
EMI source
Improper connection of
external SIO cable (Control
panel cable) Or Connection
problem with external SIO
cable connector (Lemosa
plug)
Check front panel cable and
plug (Lemosa) and replace, if
necessary
xx15
B
SIO data non plausible
Defective uC controller B
Replace PC board B
ESD or EMI problems
Check environment for ESD or
EMI source
Improper connection of
external SIO cable (Control
panel cable) Or Connection
problem with external SIO
cable connector (Lemosa
plug)
Check front panel cable and
plug (Lemosa) and replace, if
necessary
xx16
B
SIO data not received
Defective uC controller B
Replace PC board B
xx17
B
SIO cannot send
Defective uC controller B
Replace PC board B
xx20 to
xx29
and
xx50
B Safety
relevant
data
non plausible
Defective uC controller B
Replace PC board B
Defective PC board B
Replace PC board B
xx27
B
Check sum for
pressure sensor data
incorrect
Calibration does not match
Recalibrate all sensors
Содержание Centiva/5 Plus
Страница 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Страница 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Страница 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Страница 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Страница 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Страница 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Страница 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Страница 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Страница 178: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 2 9 1 Pneumatic diagram ...
Страница 180: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 4 9 2 Electrical diagram ...
Страница 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Страница 182: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 6 9 3 Component layout PC board B ...
Страница 183: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 7 9 4 Component layout PC board A ...
Страница 192: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 16 Notes ...
Страница 206: ...10 Configuration and optionalization TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 10 14 Notes ...
Страница 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Страница 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...