9 Schematics and Diagrams
TRM, Centiva/5 Plus, M1047448/CM-922026, rev. 3.50.a of 11/05
9-15
Connection
Ohmeda Com 1.0
SalTim Com 1.0
Centiva/5 provides capabilities to transfer ventilation parameter, pressure - and
flow waveform, all measured values and alarms to an external data management
system via a serial data protocol.
Centiva/5 provides two protocols on demand to a data management system:
-
Ohmeda Com 1.1
-
SalTim™ Com 1.0
Without any request from an external data management on the serial port
Centvia/5 disables the serial data transfer.
With a request from an external data management on the serial port Centvia/5
activates the Ohmeda Com 1.1 protocol as the default.
With a special request for the SalTim protocol from an external data management
on the serial port Centvia/5 activates the SalTim Com 1.0 protocol.
Connect a RS 232 data cable at the rear side of Centiva/5 at the serial data
connection (D-Sub 9 female) with the following Pin constellation:
-
Pin 2
receive data
-
Pin 3
transmit data
-
Pin 5
signal ground
Correct connection is provided with serial cable P/N BE-120116.
Centiva/5 uses the Ohmeda com 1.0 serial protocol (AESTIVA 7900 & 7100
Ventilators) as the default protocol. This allows connecting Centiva/5 to all data
management systems using this protocol.
When connecting to a /5 system use the DIS 7900 Non-Aestiva connection box.
This box provides the correct D-Sub 9 connector to connect to the Centiva/5. Do
not use the DIS 7900 AESTIVA with a D-Sub 15.
Proper communication with the DEIO Clinisoft data management system is tested.
For more details refer to information provided by the data management distributor.
Centiva/5 uses the SalTim 1.0 serial protocol upon request by CED™ software. This
allows connecting Centiva/5 to CED software systems for service and data
management purpose.
Connect the RS 232 data cable at the rear side of Centiva/5 at the serial data
connection (D-Sub 9).
For more details refer to information about CED software provided by TIM GmbH
(www.tim-gmbh.de).
9.10 Serial data transfer
Содержание Centiva/5 Plus
Страница 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Страница 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Страница 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Страница 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Страница 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Страница 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Страница 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Страница 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Страница 178: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 2 9 1 Pneumatic diagram ...
Страница 180: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 4 9 2 Electrical diagram ...
Страница 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Страница 182: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 6 9 3 Component layout PC board B ...
Страница 183: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 7 9 4 Component layout PC board A ...
Страница 192: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 16 Notes ...
Страница 206: ...10 Configuration and optionalization TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 10 14 Notes ...
Страница 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Страница 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...