10 Configuration and optionalization
TRM, Centiva/5 Plus, M1047448/CM-922026, rev.3.50.a of 11/05
10-2
Note
Centiva/5 Plus allows user specific configuration of the default machine settings
according to hospital’s needs for the following parameters:
1.
Language
2.
Date format
3.
Units of pressure
4.
Units CO2 (not applicable)
5.
Unit of ideal body weight
6.
Patient type
7.
Cross over weight
8.
Start ventilation mode
For safety reasons (and with respect to ergonomic and user interface safety theory)
all Centiva/5 Plus within one particular application or operation area (e.g. within
one CC station) should be set to the identical configuration, so that all machines
within one operation area behave identical in terms of user surface and default
start.
This configuration is not accessible to the user to avoid miss-settings of default
configuration settings. It is only accessible for authorized service personal.
10.1 Centiva/5 Plus configuration mode
SW 3.nn.p HW o/r
Centiva/5 +
Bi-Level VG
Newborn
Trend
Abort
Abort
Safe
Save
Serial number
Option key
171050101
123456
Option menu
Language
Unit Meas.
Unit Co2
Unit IBW
Patient type
Cross over weight
Start vent. Mode
English
cmH2O
mmHg
kg
35 kg
Bi-Level
Adult
Default settings
Содержание Centiva/5 Plus
Страница 1: ...GE Healthcare Centiva 5 Plus Technical Reference Manual ...
Страница 14: ...1 Introduction TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 1 8 Notes ...
Страница 20: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 6 ...
Страница 27: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 13 ...
Страница 30: ...2 Theory of operation TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 2 16 Notes ...
Страница 34: ...3 Post service check out TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 3 4 Notes ...
Страница 70: ...4 Test and calibration TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 4 36 Notes ...
Страница 154: ...7 Repair procedures TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a 7 48 Notes ...
Страница 178: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 2 9 1 Pneumatic diagram ...
Страница 180: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 4 9 2 Electrical diagram ...
Страница 181: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 5 ...
Страница 182: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 6 9 3 Component layout PC board B ...
Страница 183: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 7 9 4 Component layout PC board A ...
Страница 192: ...9 Schematics and Diagrams TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 9 16 Notes ...
Страница 206: ...10 Configuration and optionalization TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 10 14 Notes ...
Страница 212: ...11 Optional accessories TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 11 6 Notes ...
Страница 213: ...TRM Centiva 5 Plus M1047448 CM 922026 rev 3 50 a of 11 05 Notes ...