![GE AxiChrom 100 Скачать руководство пользователя страница 66](http://html.mh-extra.com/html/ge/axichrom-100/axichrom-100_operating-instructions-manual_128028066.webp)
Select whether to create a method for Column Packing, Packing Test, or the two
combined. Then make the selections in the table below.
Figure 5.1: Intelligent Packing wizard (1 of 5), ÄKTApilot.
Select “AxiChrom xxx/yyy” (xxx = column diameter, yyy = max
bed height for short and long column tube, 300 or 500 mm re-
spectively). The column size can be found on a label on the col-
umn tube near the bottom flange.
Column
Select the appropriate material. The only available option is Glass.
Column
material
Select the appropriate bed support type and material. The mate-
rial options are stainless steel and plastic. The porosity choice
(10 or 20 µm) is dependent on the bead size of the medium.
Bed support
type
Select the appropriate media type. For custom media see Creating
an Intelligent Packing method based on custom media, on
page 71.
Show
medium type
Select the appropriate BioProcess medium.
Medium
66
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
5 Operation
5.1 Intelligent Packing of the column with ÄKTAexplorer and ÄKTApilot