![GE AxiChrom 100 Скачать руководство пользователя страница 125](http://html.mh-extra.com/html/ge/axichrom-100/axichrom-100_operating-instructions-manual_128028125.webp)
3
Gently pull on the distributor to withdraw a length of the tube [405-1] from inside
the adapter rod [802].
Figure 6.12: Withdrawing a length of the tubing from inside the adapter rod.
4
Unscrew the adapter tube [405-1] from the top distributor [402] by hand.
Figure 6.13: Unscrewing the adapter tubing from the distributor.
5
Pull out the adapter tube [405-1] from the top end of the adapter rod [802]. Unscrew
the tube connection [109] from the adapter tubing [405-1]. Check the tubing and O-
rings [602a, b] for visible signs of damage. Replace if necessary.
6
Remove the scraper seal [605] and O-ring [604] by gently moving them up the dis-
tributor a bit at a time around the circumference.
The O-ring is secured within the scraper seal so they will most likely be
removed together.
Note:
7
Check the scraper seal for scratching and wear, and the O-ring for twisting and wear.
If damaged or if there is any doubt about their integrity, discard them.
8
Rinse the distributor with water and then inspect for signs of damage. If necessary
replace the distributor.
9
Check the adapter plate O-ring [127] for visible signs of damage. If damaged replace
it with a new one.
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
125
6 Maintenance and service
6.2 Performing maintenance and service