5.8
Emptying the column
WARNING
Remove the mechanical locking equipment before unpacking the
column. If it is not removed over-pressure may inadvertently be
created in the hydraulic chamber.
1
Unpack the column (see Section 5.5 Unpacking a packed column, on page 102).
2
Remove the adapter and top plate assembly (see Removing the top plate and adapter,
on page 49).
3
Release the pin at the pivot stand, if used, tilt the column and pour out medium into
a container.
4
Pump in packing liquid upflow if needed to help in pouring the medium.
Figure 5.8: Emptying the column using the pivot stand.
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
105
5 Operation
5.8 Emptying the column