Servicing the check valve in the
hydraulic chamber inlet (100
column)
1
Unscrew the check valve adapter {124} from the top bayonet [804].
2
Unscrew the check valve housing {116} from the check valve adapter {124}.
3
Fit in a new check valve [118]. The arrow on the check valve represents the flow di-
rection and shall be fitted so that it points inwards to the column.
Figure 6.28: Positioning of the check valve on AxiChrom 100 column.
CAUTION
To avoid risks caused by over-pressure the check valve shall
be fitted into the valve housing such that the direction of the
printed arrow on the valve points inwards to the column.
4
Screw the check valve adapter back into the top bayonet.
5
Screw the check valve housing back into the check valve adapter. Be careful not to
over tighten.