CAUTION
Beware when handling the sharp edges of the bayonet threads.
NOTICE
Insert the adapter into the mouth of the column at an angle,
to avoid damage to the glass tube and to avoid trapping air
under the adapter bed support.
NOTICE
Always wear gloves when handling the adapter rod to avoid
fat from unprotected fingers and scratching the sealing surface.
NOTICE
If there is slurry in the column, to avoid damage to the scraper,
make sure that there is a clear head of liquid 1-2 cm above the
top surface of the settling slurry before inserting the adapter
into the column tube.
NOTICE
Never force the adapter into the column manually. Only use
hydraulic pressure as described in this Operating Instructions
to move the adapter in the column tube.
6
Align the hole for the bayonet tool on the top plate with the furthest-right machined
mark at the top of the bayonet [804].
7
Insert one of the bayonet tools into the hole on the top plate, and the other tool into
the hole on the top bayonet (see Figure 4.5).
44
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
4 Getting started
4.6 Routine procedures