WARNING
Remove any spillage on the floor immediately to minimize the
risk for slipping accidents.
CAUTION
Removing the top plate and adapter often causes spillages and
splashing. Always use safety glasses and protective clothing.
5.6
Cleaning-in-place (CIP)
Description
Efficient cleaning methods should be developed as an integral part of the complete
process to maximize the working life of the medium. CIP removes precipitated material,
strongly bound substances and other contaminants from the medium without the need
to disassemble the column. Regular CIP between batches helps assure the desired
product quality and the expected working life cycle of the medium and equipment.
For more information about which chemicals to use for CIP, see the table on page 27.
Ensure that the media and the column withstands the chemicals used for CIP.
Note:
5.7
Sanitization
Description
Sanitization is the use of chemical agents to reduce a microbial population to an accept-
able, predetermined level.
A sanitization procedure is used when there is a need to ensure microbial reduction, for
example before changing a product or when there has been unwanted microbial con-
tamination.
Low liquid velocity (60 cm/h) is used for the whole process.
Before the sanitization method is started, check that the media withstands
1.0 M NaOH.
Note:
AxiChrom 50, 70 and 100 columns Operating instructions 28-9331-08 AG
103
5 Operation
5.5 Unpacking a packed column