Junction Box Harnesses
Symptom
Potential Causes / Solutions
It is difficult to connect the
Junction Box Harness connectors
1. Verify that the connectors are aligned with the alignment
marks on the sockets. Reference Section 4.5 and Section
4.6 for information on connecting the Junction Box
Harnesses.
2. Inspect the connectors for damage. If the connectors are
damaged, contact Galil Medical Customer Service. Do not
use a damaged Junction Box Harness.
It is difficult to loosen the Junction
Box Harness gas connection hub
The gas lines were not vented and the gas lines are still under
pressure.
WARNING. Do not apply excessive force to release the
Junction Box Harness gas connection hub. The gas lines
may still be under pressure.
•
Verify that the gas cylinders are CLOSED.
•
If the Visual-ICE MRI System is OFF, turn on the system
and wait thirty seconds for the system to automatically
vent.
•
If the Visual-ICE MRI System is ON, end the procedure and
vent the system using the automatic vent function.
•
Upon completion, close the Manual Vent Valve.
•
Contact Galil Medical Customer Service.
It is difficult to disconnect the
Junction Box Harness connectors
Reference Section 7.1 for information on disconnecting the
Junction Box Harnesses.
10.7
Mobile Connection Panel
Symptom
Potential Causes / Solutions
It is difficult to move the Mobile
Connection Panel
Verify that all Wheel Lock Levers are unlocked. Reference
The Gas Pressure Indicator LED
is not lit
1. Ensure the Junction Box Harnesses are connected.
Reference Section 4.5 and Section 4.6 for information on
connecting the Junction Box Harnesses.
2. Ensure the system is powered on.
3. Contact Galil Medical Customer Service
The Test Channel 8 button does
not respond
1. Verify that there is a needle in channel 8 and that the
locking bar is engaged.
2. Press the
Test
button for channel 8 on the Console
3. Verify that the system is powered on.
4. Verify that the Junction Box Harnesses are connected.
Reference Section 4.5 and Section 4.6 for information on
connecting the Junction Box Harnesses.
5. Contact Galil Medical Customer Service.
Содержание Visual-ICE MRI
Страница 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Страница 20: ...1 10 User Manual This page intentionally left blank...
Страница 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Страница 56: ...4 14 User Manual This page intentionally left blank...
Страница 66: ...5 10 User Manual This page intentionally left blank...
Страница 80: ...9 2 User Manual This page intentionally left blank...
Страница 118: ...12 2 User Manual This page intentionally left blank...
Страница 120: ...13 2 User Manual This page intentionally left blank...
Страница 121: ......