![Galil Visual-ICE MRI Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/galil/visual-ice-mri/visual-ice-mri_user-manual_3251268050.webp)
4-8
User Manual
1
2
3
2 4
Align gas connection hub
Turn counterclockwise to locked position
Figure 4-3. Connection of Junction Box Harness Gas Line
1
Alignment Mark
3
Ridge
2
(Red) Unlocked Position
4
(Green) Locked Position
d.
Connect the fiber optic connector on the Junction Box Harness to the fiber optic socket on the
junction box (Figure 4‑4). To make the connection, align the alignment mark (callout 1) above
the fiber optic socket with the red alignment dot (callout 2) on the fiber optic connector and
push in the connector. A clicking sound indicates a successful connection.
NOTE:
To verify the fiber optic connector is secure, pull on the check connection area identified
1
2
3
Figure 4-4. Connection of Junction Box Harness Fiber Optic Line
1
Alignment Mark
2
Red Alignment Dot
3
Check Connection Area
e.
Connect the electrical connector on the Junction Box Harness to the electrical socket on the
junction box (Figure 4‑5). To make the connection, align the alignment mark (callout 1) on the
electrical socket with the red alignment dot (callout 2) on the electrical connector and push in
the connector. A clicking sound indicates a successful connection.
Содержание Visual-ICE MRI
Страница 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Страница 20: ...1 10 User Manual This page intentionally left blank...
Страница 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Страница 56: ...4 14 User Manual This page intentionally left blank...
Страница 66: ...5 10 User Manual This page intentionally left blank...
Страница 80: ...9 2 User Manual This page intentionally left blank...
Страница 118: ...12 2 User Manual This page intentionally left blank...
Страница 120: ...13 2 User Manual This page intentionally left blank...
Страница 121: ......