Message
Reason for Appearance / Solutions
80-31 System Error
•
Gas cylinder shut off valve is not open enough
to provide adequate flow. If necessary, open the
valve approximately another half turn.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
Reason for Appearance:
Gas flow from the gas cylinder was
insufficient. To improve gas flow, the
cylinder valve should be opened further.
Solutions:
Check that the gas cylinder valve is
sufficiently opened to allow gas flow
(reference Section 4.7). If necessary, open
the valve approximately another half turn.
80-32 System Error
•
Channel X is defective. Choose another channel.
At the conclusion of the procedure, contact Galil
Medical Customer Service.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
Reason for Appearance:
A defective solenoid on Channel X was
detected; another channel should be
selected.
Solutions:
1.
Select a different channel.
2. After completing the procedure,
contact Galil Medical Customer
Service.
80-34 System Error
•
Fan X failure. At the conclusion of the procedure,
contact Galil Medical Customer Service.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
Reason for Appearance:
Fan X was identified as not operating.
Solutions:
After completing the procedure, contact
Galil Medical Customer Service.
80-35 System Error
•
The gas flow rate in channel XX exceeds the
system operation requirements and may impact
performance. Reduce the number of active
needles.
•
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o
Reason for Appearance:
The calculated flow rate for a given
channel exceeded the recommended flow
rate for a channel. The number of active
needles should be reduced.
Solutions:
1. Reduce the number of active needles.
2. Contact Galil Medical Customer
Service for further guidance.
Содержание Visual-ICE MRI
Страница 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Страница 20: ...1 10 User Manual This page intentionally left blank...
Страница 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Страница 56: ...4 14 User Manual This page intentionally left blank...
Страница 66: ...5 10 User Manual This page intentionally left blank...
Страница 80: ...9 2 User Manual This page intentionally left blank...
Страница 118: ...12 2 User Manual This page intentionally left blank...
Страница 120: ...13 2 User Manual This page intentionally left blank...
Страница 121: ......