9 SYSTEM CARE and MAINTENANCE
9.1 Cleaning
9.1.1 Cleaning the Visual-ICE MRI System
Clean the Visual-ICE™ MRI Cryoablation System after each use by following the steps below.
1. Clean the Touch Screen Monitor when the Visual-ICE MRI Cryoablation System is powered OFF.
•
Wipe the screen gently with a damp gauze wipe.
•
Use water or isopropyl alcohol only.
•
Do not use any other cleaning agent such as Betadine or bleach solution.
2. Clean the Console, Mobile Connection Panel, and Junction Box Harnesses by wiping each with a
damp gauze wipe.
•
Use soap and water or isopropyl alcohol cleaning solutions.
•
Do not use cleaning agents such as Betadine or bleach solution.
•
Do not allow water or other liquid to drip or seep inside the needle connection ports of the
Needle Interface on the Mobile Connection Panel. The needle connection ports must remain
completely dry at all times.
3. Ensure that cleaned surfaces are dry prior to closing or turning on the system.
CAUTION.
Do not set food, drinks, or other objects on top of the Visual-ICE MRI Cryoablation
System. Doing so may damage the system.
9.2 Service and Preventive Maintenance
Only trained and authorized service engineers are permitted to conduct service and system
preventive maintenance on the system. Preventive maintenance on the Visual-ICE MRI Cryoablation
System is performed every two years.
CAUTION.
Do not attempt to service the Visual-ICE MRI Cryoablation System. If service is required,
contact Galil Medical Customer Service.
The Visual-ICE MRI Cryoablation System will display a reminder on the screen approximately one
month prior to when the preventive maintenance is due. If the reminder message appears and the
preventive maintenance has not yet been scheduled, contact your Galil Medical authorized service
representative to schedule the service.
9.3 System Disposal
In compliance with Waste Framework Directives, when a Visual-ICE MRI System is decommissioned,
contact Galil Medical Customer Service to schedule system replacement, refurbishing or disposal.
9.4 Technical Information
If additional technical information is required, place a request with a Galil Medical qualified service
engineer.
Содержание Visual-ICE MRI
Страница 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Страница 20: ...1 10 User Manual This page intentionally left blank...
Страница 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Страница 56: ...4 14 User Manual This page intentionally left blank...
Страница 66: ...5 10 User Manual This page intentionally left blank...
Страница 80: ...9 2 User Manual This page intentionally left blank...
Страница 118: ...12 2 User Manual This page intentionally left blank...
Страница 120: ...13 2 User Manual This page intentionally left blank...
Страница 121: ......