![Galil Visual-ICE MRI Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/galil/visual-ice-mri/visual-ice-mri_user-manual_3251268043.webp)
Visual-ICE
™
MRI Cryoablation System
4-1
4 SYSTEM SETUP
depicts the order and steps of Visual-ICE™ MRI Cryoablation System setup. This section
describes each step in detail.
Table 4-1. System Setup Flow
1.
System
Set-Up
•
Confirm availability of gas, needles, and accessories
•
Position Visual-ICE MRI Mobile Connection Panel in
the magnet room and lock all wheels
•
Position Visual-ICE MRI Console in the control room
and lock brake
•
Ensure Manual Vent Valve is closed and Argon
Shut‑off Valve is ON
•
Power Visual‑ICE MRI Console ON
•
Login
2.
Connect Gas
Cylinders
•
Connect the helium gas supply line to the Console
and then connect the gas supply line to the helium
cylinder
•
Connect the argon gas supply line to the Console
and then connect the gas supply line to the argon
cylinder
•
Ensure the safety cables are connected
3.
Connect
Junction Box
Harnesses
•
Connect Junction Box Harness in the control room
•
Connect Junction Box Harness in the magnet room
•
Ensure the safety cables are connected
NOTE:
Drape harnesses to avoid trip hazards
4.
Open Gas
Cylinder Valves
•
OPEN valve on helium, then argon
•
Verify gas pressure is within working limits for the
cryoablation procedure (reference Table 4-2)
4.1 Preparation for Use
Prior to using the Visual‑ICE MRI System, inspect the Console, Mobile Connection Panel, power
cord, brakes, safety cables, gas supply lines, gas connections, Junction Box Harnesses, and the
monitor touch screen to ensure they are not damaged.
If any of the components are damaged, contact Galil Medical Customer Service.
CAUTION.
If condensation is present on the system, allow the system to dry completely for 12 hours
prior to powering up the system. Powering up the system with condensation could result in permanent
damage to the electrical boards, causing the system to be inoperable.
Содержание Visual-ICE MRI
Страница 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Страница 20: ...1 10 User Manual This page intentionally left blank...
Страница 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Страница 56: ...4 14 User Manual This page intentionally left blank...
Страница 66: ...5 10 User Manual This page intentionally left blank...
Страница 80: ...9 2 User Manual This page intentionally left blank...
Страница 118: ...12 2 User Manual This page intentionally left blank...
Страница 120: ...13 2 User Manual This page intentionally left blank...
Страница 121: ......