MRI Cryoablation System
10-3
10.2
Electronics, Electrical and User Error Related Issues
Symptom
Potential Causes / Solutions
System does not turn ON (i.e., the
fan is not running) or power is lost
during procedure
1. The power control knob on the front panel of the
Console or power switch on the rear panel is OFF
(Figure 2-2 and Figure 2-3). Turn the power ON.
2. The power cable to the Visual-ICE MRI Console is
disconnected from the power outlet or from the rear
panel of the Console. Connect the power cable to the
Visual-ICE System, checking that the power cable is fully
seated. Connect the power cable to the power outlet.
3. There is no power at the wall outlet. Ensure the power
outlet is switched ON. Call the hospital Biomedical
Engineer for assistance, if needed.
4. A fuse may have blown. Spare fuses are located in the
power cord inlet on the system (Figure 2-3). Reference
Section 10.2.1 for information on how to replace fuses in
the system.
Channel or needle is not
recognized by the system and not
available for use
1. Verify that the Junction Box Harnesses are connected.
Reference Section 4.5 and Section 4.6 for information
on connecting the Junction Box Harnesses.
2. Check the corresponding Channel Locking Bar and
ensure it is fully in the locked position.
3. At least one needle must be inserted into a channel for
the channel to be available for use.
4. The channel may be defective. Do not use this channel.
Move the needle(s) to a different channel. Conduct the
Needle Integrity and Functionality Test again.
USB flash drive is not functioning
Or
The USB flash drive is not
recognized by the system
1.
The USB flash drive is not properly connected to the
USB port. Remove the USB flash drive from the USB
port on the Console. Wait a few seconds and reconnect
the USB flash drive to the USB port.
2.
If the problem persists, try using a different Galil Medical
supplied USB flash drive.
3.
The USB flash drive is faulty. Replace the USB flash
drive with a new Galil Medical supplied USB flash drive.
The
Login Screen
appeared
after the system was left idle for
more than 2 hours while on the
Procedure Screen
Enter the appropriate password to return to the
Procedure
Screen.
Содержание Visual-ICE MRI
Страница 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Страница 20: ...1 10 User Manual This page intentionally left blank...
Страница 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Страница 56: ...4 14 User Manual This page intentionally left blank...
Страница 66: ...5 10 User Manual This page intentionally left blank...
Страница 80: ...9 2 User Manual This page intentionally left blank...
Страница 118: ...12 2 User Manual This page intentionally left blank...
Страница 120: ...13 2 User Manual This page intentionally left blank...
Страница 121: ......