NOTE:
Gas cylinder connections have left‑handed threads.
2
1
Figure 4-2. Gas Cylinder Set-up
1
Gauge assembly adapter
2
Cylinder valve
8.
Repeat the procedure outlined in Steps 4 through 7 to connect the argon gas to the Console,
using the argon gas supply line.
4.5 Connecting the Junction Box Harness in the Control Room
1.
Position one end of the Junction Box Harness next to the junction box in the control room.
a.
Secure the safety cable at the end of the gas line to the safety cable ring on the
junction box.
b.
Remove the protective covers from the Junction Box Harness connectors and junction box
sockets.
c.
Connect the gas connection hub on the Junction Box Harness to the gas connection socket
on the junction box (Figure 4‑3). To make the connection, align the ridge (callout 3) and red
unlocked position marker (callout 2) on the gas connection hub with the alignment mark
(callout 1) above the gas connection socket, push in the gas connection hub, and turn the gas
connection hub counterclockwise to the locked position (callout 4). A clicking sound indicates
a successful connection.
Содержание Visual-ICE MRI
Страница 1: ...M R I C R Y O A B L A T I O N S Y S T E M RX Only User Manual...
Страница 20: ...1 10 User Manual This page intentionally left blank...
Страница 42: ...3 12 User Manual Screen 3 12 Successful Upload Download...
Страница 56: ...4 14 User Manual This page intentionally left blank...
Страница 66: ...5 10 User Manual This page intentionally left blank...
Страница 80: ...9 2 User Manual This page intentionally left blank...
Страница 118: ...12 2 User Manual This page intentionally left blank...
Страница 120: ...13 2 User Manual This page intentionally left blank...
Страница 121: ......