97
Maintenance Menu
7.
Diagnostic Test Function
The FSM-60S has a built in diagnostic test feature that allows the operator to perform a
simple one step evaluation of splicer performance covering several different critical
variables. Perform this function in the event of splicer operation trouble.
Operation Procedure
(1) Select the [Diagnostic Test] in the [Maintenance Menu 2] and execute [Diagnostic
Test]. The following checks will be made.
Check
Item
Description
1 LED
Check
Measures and adjusts the brightness of the
illumination LED.
2
Motor Check
Check the Motor Limit Sensor.
3 Dust
Check
Checks the optical path for dust or dirt and judges
whether they disturb fiber observation. If
contamination exists, this function indicates the
location.
4 Motor Calibration Automatically calibrates the 6 Motor Speeds.
5 Arc
Calibration
Automatically calibrates the arc power factor and
fiber splicing position.
6
I/O Port Check
Checks for normal operation of the input and output
terminals of the internal circuit.
7
Memory Check
Checks the memory of the internal circuit.
(2) Upon completion of all checks and adjustments, a list of results is displayed.
If the dust check result is not good, clean the objective lenses. In the case that
cleaning cannot eliminate contamination, there is a possibility that the
contamination may have entered the inside of the optical path. Please contact your
distributor for additional instructions.
The dust check and Motor Calibration functions exist as independent instructions in
[Maintenance Menu 2]. It is possible to execute them independently.
• Before the start of the test, remove the fibers from the splicer.
• When item 2 (dust check) is completed, prepare and load the fibers into the
splicer and press Enter key.