104
Error Message List
Error
Message
Reason Solution
• The fiber is set too far
back and does not reach
the splice point.
• Press RESET key, re-position the fiber again
with the end-face closer to the electrodes.
ZL Motor
Overrun
(Forward)
• The fiber is not set
correctly at the bottom
of the V-groove. The
fiber is not located in
the Camera’s field of
view.
• Press RESET key, and set the fiber again to
seat it correctly at the bottom of the
V-groove.
ZR Motor
Overrun
(Forward)
• The cleave length (bare
fiber part) is too short.
• Confirm the setting position of the stripped
fiber end on the fiber cleaver. Check the
cleave length.
ZL Motor
Overrun
(Backward)
The taper speed or taper
time is set too high.
Adjust the taper parameters in the splice
mode.
ZR Motor
Overrun
(Backward)
Only occurs in manual motor operation.
X motor
Overrun
(Forward)
X motor
Overrun
(Backward)
Y Motor
Overrun
(Forward)
Y Motor
Overrun
(Backward)
The fiber is not set
correctly at the bottom
of the V-groove. The
fiber is offset too far and
it exceeds the X or Y
motor range.
Press RESET key and re-position the fiber
again to seat it correctly at the bottom of the
V-groove.