44
Basic Operation
(3)
After completion of fiber alignment, arc discharge is performed to splice the fibers.
(4)
Estimated splice loss is displayed upon completion of splicing. Splice loss is
affected by certain factors stated in the next page. These factors are taken into
account to calculate, or estimate, splice loss. The calculation is based on certain
dimensional parameters, such as MFD.
If either the cleave angle measured or the estimated splice loss exceeds its set
threshold, an error message is displayed. If the spliced fiber is detected as abnormal,
such as "Fat", "Thin" or "Bubble", an error message is displayed.
If no error message is displayed but the splice looks poor by visual inspection
through the monitor, it is strongly recommended to repeat the splice from the
beginning.
Splice loss may be improved in some cases by additional arc discharges. Press ARC
key for an additional arc discharge (re-arc). Splice loss estimate and splice check are
performed again.
• Splice loss may be worsened in some cases by additional arc discharges (re-arcs).
Additional arc discharge can be set to "disabled", or limited to the number of
additional arcs. See [Splice Settings] page 84 for detail.