101
Maintenance Menu
10.
Motor Calibration
Motors were adjusted at the factory before shipping. However, settings could change due
to various reasons. This function automatically calibrates the speed of all six motors.
Operation Procedure
(1) Select the [Motor Calibration] in the [Maintenance Menu 2].
(2) Load prepared fibers in the splicer and press Enter key.
(3) Speeds for all motors are automatically calibrated. Upon completion, [Maintenance
Menu 2] is displayed.
11.
Motor Drive
The six motors incorporated in the splicer can be manually operated individually. In the
course of splicing, the motors can also be operated by calling this menu in the [PAUSE1],
[PAUSE2] or [FINISH] state.
(1) Select the [Motor Drive].
(2) Pressing Menu key changes motor selection. The name of the selected motor is
displayed in the upper section of the screen.
(3) Press Up/Down Arrow key to drive the motor in the desired direction.
Motor
Up Arrow key
Down Arrow key
ZL/ZR Forward.
Backward.
X/Y Upward.
Downward.
Focus X
Focus Y
Lens moves nearer fiber.
Lens moves away from fiber.
• When the motor reaches the limit of the operating range, the buzzer sounds and the
motor stops. Press the opposite arrow key to reverse and move the motor again.
• Display messages can be erased by pressing Enter key. The message can be displayed
by pressing Enter key again.
• If you finish this operation by pressing RESET key, the motor will return to the reset
position. To finish the operation with the motor held in the adjusted position, press
Escape key.
• If the motor is moved too much with respect to the spliced fiber, the fiber may break.
• Perform this function when "Fat" or "Thin" error has occurred, or fiber
aligning or focusing is taking too much time.