Basic Operation
Symptom Cause
Remedy
Arc power not
adequate
Perform [Arc Calibration].
Thin
Some arc parameters
not adequate
Adjust [Prefuse Power], [Prefuse Time] or
[Overlap].
Line
Some arc parameters
not adequate
Adjust [Prefuse Power], [Prefuse Time] or
[Overlap].
Storing splicing results
Splicing results is stored in memory.
The splicing result is automatically stored in memory when SET key or RESET key is
pressed upon completion of the splice at the [Finish] screen, or when the wind
protector is opened upon completion of the splice at the [Finish] screen.
Once a certain comment is inputted, the same comment is inputted into subsequent
splice results. To change comments, see [Splice Memory Comment].
• A vertical line sometimes appears at the splice point when MM fibers, or
dissimilar fibers (different diameters) are spliced. This does not affect splice
quality, such as splice loss or tensile strength.
• After the 2000th result is stored, 2001st splice result is written over 1st result.