99
Maintenance Menu
9.
Arc Calibration
Atmospheric conditions such as temperature, humidity, and pressure are constantly
changing, which creates variability in the arc temperature. This splicer is equipped with
temperature and pressure sensors that are used in a constant feedback monitoring control
system to regulate the arc power at a constant level.
Changes in arc power due to electrode wear and glass adhesion cannot be corrected
automatically. Also, the center position of arc discharge sometimes shifts to the left or
right. In this case, the fiber splicing position has to be shifted in relation to the arc
discharge center. It is necessary to perform an arc power calibration to eliminate both of
these issues.
Operation procedure
(1) Select [Arc Calibration] in [Maintenance Menu 2] to display Arc Calibration screen.
(2) Set prepared fibers in the splicer.
(3) The following is performed after pressing
Enter key.
1. Calculating the Arc center
Arc discharge is performed without fibers in the view screen to detect the center of
arc discharge and adjust the gapset position.
2. Cleaning arc discharge
The left and right fibers move forward. The splicer performs the cleaning arc
discharge.
3. Gap setting
The left and right fibers further move forward and stop at the specified gap setting
position.
4. Arc discharge
The splicer performs an arc discharge without stuffing the fibers together. . The gap
between the left and right fibers is melted back by the arc discharge.
5. Measurement result
After the arc discharge, the melt-back amounts of the left and right fibers are
measured by image processing circuits in the splicer.
• Arc calibration is performed automatically using [AUTO] mode only. So arc
calibration does not have to be performed when splicing in this mode.
• Performing the [Arc Calibration] function changes the arc power “factor”
value. The factor value is used in the algorithm program for all splicing. The
arc power value will not change in the splice modes.
• Use standard SM or DS or MM fiber for arc calibration.
• Use well prepared fibers for arc calibration. Dust on the fiber surface affects
arc calibration.