Appendix A
Optics Controls
This table explains how to use each of the optics controls, which are used during manual inspections.
See
on page 60 for more information.
Control
Imaging Methods
Description
Excitation
●
Fluorescence
The
Excitation
list provides you with three different light
wavelengths, including the standard package
Ultraviolet 290
nm
,
Visible Fluorescence 485 nm
, and
Visible Fluorescence
578 nm
.
Available
Excitation
list depends on the filter modules you
purchased.
Mode
●
UV
The
Mode
list provides you with two UV imaging options,
including Absorption and Fluorescence.
UV absorption imaging is available to new ROCK IMAGER 1000
models only.
Objective
●
SLP Visible
●
UV
●
Fluorescence
●
FRAP
●
SONICC
To change the
Objective
values, click the corresponding list, and
then select the desired value as necessary
Zoom
●
Visible
●
SLP Visible
●
UV
The
Zoom
field is generated by the current zoom setting and is
related to the FOV field. These fields allow you to zoom in on
drop locations and regions of interest within a drop or well.
To change the Zoom values, type a new value in the
corresponding box, and press
Enter
on your keyboard.
Alternatively, click the left and right arrows to decrease or
increase the values.