High-Pressure Modular Controller
Configure the Driver
51
Configure the Driver
Turn on the CPS, SPLT and/or valve accessories from the Instrument Setup
menu (
Setup>Instrument Setup
). These accessories require the correct
pneumatic connections and electrical connections.
See their respective
instruction sheets for more information.
Touch the
External 24V
tab to see or set
the state of the external drivers.
Drivers are identified in the External 24V menu and on the rear panel of the
Product as DRV1, DRV2, DRV3, DRV4 (see Figure 12). Each solenoid uses two
wires, one wire goes into each connector.
DRV1 uses the two top left inputs
DRV2 uses the two top right inputs
DRV3 uses the two bottom left inputs
DRV4 uses the two bottom right inputs
Driver 2
Driver 4
Driver 1
Driver 3
huo031.eps
Figure 12. Driver Location
Several of the drivers support specific accessories such as the CPS, SPLT and
Abort/Isolation valves for System Mode.
DRV1: CPS Valve, or SPLT
DRV2: CPS Valve
DRV3: CPS LED, or External Isolation Valve
DRV4: External Abort Valve
See the associated appendices for additional information on these accessories.
The External 24 V menu (
Setup>Instrument Setup
>
External 24V)
also shows
the driver states for any auxiliary chassis when working in a system with multiple
chassis.