High-Pressure Modular Controller
PCM Installation
11
PCM Installation
Depending upon how the Product is ordered, the PCM may come installed in the
unit or be delivered in a separate box.
To install the PCM:
W
Caution
To prevent damage to the sensors inside the modules, do not
drop the modules.
1. Remove the PCM from its shipping box.
2. Remove any protective plastic covers from the PCM pressure ports.
3. Open the Module Bay as described in the
Access the Module Bay
section.
4. The PCM location in the Module Bay is on the far right. Align the track on the
bottom of the PCM with the PCM slot inside the Module Bay. See Figure 3.
5. Slide the PCM into place.
6. Tighten the two bolts on the front of the PCM using a 6 mm hex head wrench.
The recommended torque is 10-14 N ꞏ m (7 - 10 lbf ꞏ ft).
W
Caution
To prevent damage to the internal manifold, do not over torque.
7. Close and latch the front panel.
ifb021.eps
Figure 3. PCM Installation
Barometer Module
For absolute measurements with gauge mode modules, a barometer module
(PM200-A100K, PM500-A120K, PM500-BA120K, PM500-A200K, or BRM600-
BA100K) must be installed, see
Atmosphere
. The module can also be selected
as an Autozero reference, see
Autozero - Absolute Mode
.