8270A/8370A
Operators Manual
58
Remote Interface
The Remote Interface Diagnostics menu provides information on remote
communication by monitoring the input and output communication. The menu
contains independent diagnostics for RS-232, USB, Ethernet, GPIB, and the
System Bus communications.
RS-232
– Shows the Receive and Transmit data for this interface.
USB
– Shows the Receive and Transmit data for this interface.
Ethernet
– Shows the Receive and Transmit data for this interface.
GPIB
– Shows the Receive and Transmit data for this interface.
System Bus
– Shows the subsequent data for this interface.
Receive
Transmit
Fault
Troubleshooting
Table 19 explains minor troubleshooting issues. For issues outside of the scope of this
section, the Product may require service. See
How to Contact Fluke Calibration.
Table 19. Troubleshooting
Problem
Probable Cause
Action
Electrical Problems
Product
does not
turn on
Not plugged in
Verify Product is plugged in and power is available.
Power not available
Check the fuse per instruction in this manual.
Fuse blown
Verify the main power switch is ON.
Check the fan. If the fan is on or solenoid click is heard, see
the “Display does not turn on” problem.
If the fan is not on, have the power supply serviced. Verify
internal power connections.
Display
does not
turn on
Chassis power issue
Verify that power is supplied, see above.
Screen saver is
activated
Verify power to front panel.
and
should be lit.
Power plug to front
panel is loose
Verify the
is in the ON state.
Display panel failure
If
is yellow, then push it to turn on display.
Open front panel and verify the power plug is connected to
the front panel.
PCM not initializing
Remove PCM and attempt to reboot.