8270A/8370A
Operators Manual
20
Table 5. Instrument Settings Menu (cont.)
Tab
Description
Change
Password
The integrity of Product calibration is protected by a security password that must be
entered before new calibration constants can be saved to non-volatile memory. The
password also protects the ability to set the date for the internal real-time clock. If
the password has not been entered, the Product is secured.
Once the password is entered, the Product is unsecured. The Product secures itself
when the Setup menus are closed or the product is reset. The Product can be
unsecured at any time over the remote interface with the CAL_SECURE command
and by entering the password. The Product prompts for the password to unsecure
the Product before it can accept new values to be eventually secured. The
password contains 1 to 8 digits and is factory set to
8270
.
To change the password locally:
1. Touch
Setup Menu>Instrument Setup>Instrument Settings>Change
Password
. The Product prompts for the current password.
2. Use the numeric keypad to enter the current password.
To change the password over the remote interface use the CAL_PASSWD
command.
Note
If the new password is lost, contact Fluke Calibration
Customer Service. Fluke Calibration will supply a new
password.
External 24 V
Menu
Selects which driver port, on the rear of the Product, is in use.
External Valves
Use this tab to choose if a Contamination Prevention System (CPS) , Isolation Valve,
Abort Valve, Supply Isolation Valve, or Self-Purging Liquid Trap (SPLT) is installed.
See the corresponding Instructions sheet for the appropriate accessory
for more
information.
Uncertainty
Menu
Use this menu to view or change uncertainty parameters. Editable parameters are:
Head Height Uncertainty
Include Control Uncertainty
Additional Uncertainty Component 1
Additional Uncertainty Component 2
Show Uncertainty
See the Technical Note
Guide to Determining Pressure Measurement Uncertainty
for PM200, PM500, and PM600 Modules
located at
http://www.Flukecal.com
.
About this
Instrument
This page of the menu is for information purposes only. The information is useful for
the user and any technician that might work on the Product. The information on the
page includes:
Model Number
Serial Number
Revision (of the Firmware)
User Interface
Restore Factory
Default
To restore the Product to factory settings, touch this tab and then
OK
.