High-Pressure Modular Controller
Tasks
49
DUT
–
Enter the DUT uncertainty details.
If Program Type is set to Pressure the menu choices are:
Use DUT Tolerance: Yes/No
- Toggle this to
On
so that the controller goes
to the next program step without prompting the user.
Threshold Tolerance
– Sets the threshold tolerance value.
Relative Tolerance –
Sets the relative tolerance value.
Combination Method - Great of /Addition
Combine the Threshold Tolerance
and
Relative Tolerance
. At each
pressure calculate, the threshold and the relative and use whichever is
greater or as addition.
If Program Type is set to mA or DCV the menu choices are:
DUT Range (0%)
- Configure the DUT lower range
DUT Range (100%) -
Configure the DUT upper range
Tolerance Method:
% Reading
- Use reading percent as tolerance method.
% Span
- Use span percent as tolerance method.
Unit
- Use Unit (nominal mA) as tolerance method.
Allowable Error:
Configure the acceptable error, associate with Tolerance
Method.
Closed Loop: ON / OFF
For a fully-automated test where the program reads the DUT mA measurement
and then continues, set
Closed Loop
to
ON
. For manual operation, set
Closed
Loop
to
OFF
.