PROGram:REPort:REFerence <boolean>|DEFault
PROGram:REPort:REFerence?
(-P models only) Select the reference source for program reports.
<boolean> is 0 (internal control sensor) or 1(external reference probe, default).
This command is conditionally password protected.
PROGram:REPort:TOLerance <numeric_value>|DEFault
PROGram:REPort:TOLerance?
(-P models only) Set the pass tolerance for program reports.
<numeric_value> is the tolerance number. Range is 0.001 to 50. Default is 1.
READ?
(-P models only) Return the reference probe temperature and DUT sensor
reading.
The response is temperature in
°
C or
°
F and depends on the temperature unit
setting.
SENSe1:CALibration:PARameter<index> <numeric_value>
SENSe1:CALibration:PARameter<index>?
(-P models only) Set a calibration parameter for the Input Module PRT/RTD
inputs.
<index> selects the calibration parameter; 1: REF1C0; 2: REF1C100.
<numeric_value> is the value to which the calibration parameter is set.
Calibration parameters should only be changed by a qualified technician as part
of a calibration procedure.
This command is password protected.
SENSe1:DATA?
(-P models only) Return the reference probe resistance in ohms.
SENSe2:CALibration:PARameter<index> <numeric_value>
SENSe2:CALibration:PARameter<index>?
(-P models only) Set a calibration parameter for the Input Module thermocouple
and transmitter inputs.
<index> specifies the calibration parameter; 1: TCC0; 2: TCC100; 3: TCCRJ; 4:
mAC4; 6: mAC22.
<numeric_value> is the value to which the calibration parameter is set.
Calibration parameters should only be changed by a qualified technician as part
of a calibration procedure.
This command is password protected.