6109A/7109A
Operators Manual
36
•
Stability Beep controls whether the beeper sounds when the control
indicator changes to Stable.
•
Control Sensor (7109A-P and 6109A-P only) selects which
temperature sensor is used for temperature control. This feature is
useful to improve the temperature accuracy. The usual configuration
(Internal) uses the internal PRT. When set to Reference, the Product
heats or cools the tank so that the reference probe temperature
matches the setpoint.
(Setup > Operation)
°
C/
°
F
Select the temperature unit.
Fahrenheit is only available in some regions of the world.
(Setup)
Instrument
View or change configuration and user preferences.
•
Language selects the language for the user interface. Options are
English, French, German, Portuguese, Spanish, Russian, Simplified
Chinese, Japanese, and Korean.
•
Time Format selects the format for the time shown on the display.
Options are 24-hour and 12-hour format.
•
Time changes the clock’s time.
•
Date Format selects the format of dates shown on the display. Options
are YYYY-MM-DD (default), MM-DD-YYYY, DD.MM.YYYY,
DD/MM/YYYY, and YYYY/MM/DD.
•
Date changes the clock’s date.
•
Decimal selects the decimal character used in numbers shown on the
display. Options are period and comma.
(Setup > Instrument)
English
Change the UI language to English.
(Setup > Instrument)
Remote
Change settings related to the remote interface.
•
Termination selects the line termination characters for transmitted
messages. The options are CR, LF, and CR/LF.
•
Serial Baud Rate selects the bit rate for the RS-232 port.
•
Serial Monitor turns on automatic transmission of temperature readings.
The fluid temperature is sent at intervals of one second. Temperature is in
°
C or
°
F and depends on the temperature unit setting. Serial Monitor is
always OFF after power on.
(Setup > Instrument)
More...
Access additional instrument settings.