Portable Calibration Baths
Commands
45
*ESR?
Return and clear the IEEE-488.2 Standard Event Status Register.
The response is a decimal number that ranges from 0 to 255. The definition of
the bits are shown in Table 5.
Table 5. *ESR Bits
Bit Name
Definition
0 OPC
Operation
complete
1
RQC
Request control (not used)
2
QYE
Query error flag (not used)
3
DDE
Device-specific error flag
4
EXE
Execution error flag
5
CME
Command error flag
6
URQ
User request (not used)
7 PON
Power
on
*IDN?
Return product information. The response contains the manufacturer, model
number, serial number, and firmware version numbers.
*OPC
*OPC?
Operation complete. Since all commands are sequential, this command is not
necessary for synchronization.
*OPT?
Return the product configuration.
The response is 1 if the Product is equipped with the Input Module (7109A-P or
6109A-P only). It is 0 otherwise.
*RST
Reset operational settings.
•
Stop temperature control
•
Set the setpoint to 25
°
C
•
Set Control Sensor to Internal
•
Set Ramp Enable to OFF
•
Set Stability Beep to ON
•
Set Stability Window to 0.05
°
C
•
Stop program execution