CALCulate1:CONVersion:PARameter:VALue <parameter>,<n>
CALCulate1:CONVersion:PARameter:VALue? <parameter>
(-P models only) Set the value of the given reference probe characterization
coefficient.
<parameter> is the name of the characterization coefficient.
<n> is the value the characterization coefficient.
This command is conditionally password protected.
CALCulate1:CONVersion:PROGram
(-P models only) Program the reference probe memory with the current PRT
type, characterization coefficients, serial number, and calibration date.
This command is password protected.
CALCulate1:CONVersion:SNUMber <string>
CALCulate1:CONVersion:SNUMber?
(-P models only) Set the reference probe serial number.
<string> is an alphanumeric serial number or identification string that contains up
to10 characters, enclosed in quotation marks.
This command is conditionally password protected.
CALCulate1:CONVersion:STATe?
(-P models only) Return the connection state of the reference probe memory.
The response is 1 if the Product and reference probe memory are linked or 0
otherwise.
CALCulate1:CONVersion:TEST? <numeric_value>
(-P models only) Return temperature calculated from the given resistance that
uses the current reference PRT type and characterization coefficients.
<numeric_value> is the simulated PRT resistance in ohms.
The response is temperature in
°
C or
°
F and depends on the temperature unit
setting.
CALCulate1:DATA?
(-P models only) Return the reference probe temperature.
The response is temperature in °C or °F and depends on the temperature unit
setting, unless the PRT type is RES or NONE.
CALCulate2:AVERage[:AVERage]?
(-P models only) Return the moving average of the DUT sensor readings.
The response is a temperature value in °C or °F and depends on the temperature
unit setting.