6109A/7109A
Operators Manual
58
SOURce:LCONstants:DERivative <numeric_value>
SOURce:LCONstants:DERivative?
Set the derivative loop constant for temperature control.
Control parameters affect the temperature stability of the Product and should
only be changed by a qualified personnel.
This command is password protected.
SOURce:LCONstants:INTegral <numeric_value>
SOURce:LCONstants:INTegral?
Set the integral loop constant for temperature control.
Control parameters affect the temperature stability of the Product and should
only be changed by a qualified personnel.
This command is password protected.
SOURce:LCONstants:PBANd <numeric_value>
SOURce:LCONstants:PBANd?
Set the proportional band loop constant for temperature control.
Control parameters affect the temperature stability of the Product and should
only be changed by a qualified personnel.
This command is password protected.
SOURce:RAMP:RATE <numeric_value>
SOURce:RAMP:RATE?
Set Ramp Rate.
<numeric_value> is the ramp rate value in
°
C per minute or
°
F per minute.
The ramp function must be enabled for Ramp Rate to apply.
SOURce:RAMP:ENABle <boolean>
SOURce:RAMP:ENABle?
Set Ramp Enable.
<boolean> is 1(ON) or 0 (OFF).
When Ramp Enable is ON, the heating or cooling rate is reduced according to
the Ramp Rate setting. When Ramp Enable is OFF, the Product will heat or cool
as fast as possible to a new setpoint.