![Fluke 6109A Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/fluke/6109a/6109a_operators-manual_2310363042.webp)
6109A/7109A
Operators Manual
38
Security determines which menus and functions require the password.
Regardless of the Security setting, these menus are protected:
•
Setup > Instrument > More > Control
•
Setup > Instrument > More > Password
•
Setup > Instrument > More > Service > System Status > Update Firmware
•
Setup > Instrument > More > Service > Bath Calibration
•
Setup > Instrument > More > Service > Input Calibration (-P models)
•
Setup > Instrument > More > Memory > Clear Settings
•
Setup > Instrument > More > Memory > Clear Data
•
Setup > Probe > Reference Probe > Program Probe (-P models)
When Security is ON, these additional menus are protected:
•
Program > Select > Edit
•
Program > Select > Options
•
Program
>
New
•
Program
>
Delete
•
Program > Reports > Delete
•
Setup > Operation
•
Setup > Instrument
•
Setup > Probe > Reference Probe (-P models)
•
Setup > Probe > DUT (-P models)
(Setup > Instrument > More > Password)
Default
Change the password back to the factory default (1234).
(Setup > Instrument > More)
Service
View or change service related parameters.
Product information is shown (Model, Serial Number, Firmware Version,
Bath Calibration date, and Input Calibration date (-P models)).
(Setup > Instrument > More > Service)
System Status
View information about the condition of the Product.
The System Status screen shows the results of the latest self-check. A
self-check is performed when the Product is switched on.
(Setup > Instrument > More > Service > System Status)
Self Check
Repeat the system self-check.
Subsystems and components are tested. When completed, the System
Status screen shows updated results.
(Setup > Instrument > More > Service > System Status)
Diagnostics
View data about the Product operation.
The diagnostics indicators can be helpful when the Product is serviced
and for troubleshooting issues.
www.
.com
1.800.561.8187