Table 7. Program Variables
Setting Name
Description
Cycles
CYCL
Number of cycles. Range is 1 to 99. Default is 1.
Order
ORD
Setpoint order. Range is 0 (linear, default) or 1 (up/down).
Dwell Time
DWEL
Dwell time in minutes. Range is 1 to 900. Default is 10.
Ramp Rate
RRAT
Heating or cooling rate in
°
C per minute or
°
F per minute.
Ramp Enable
RREN
Turn ramp rate on or off. Range is 0 (off, default) or 1(on).
Points
POIN
Number of setpoints, 1 to 8. Default is 2.
Setpoint 1
SPO1
Setpoint temperature in
°
C or
°
F. Default is 40
°
C or 104
°
F.
Setpoint 2
SPO2
Setpoint temperature in
°
C or
°
F. Default is 40
°
C or 104
°
F.
Setpoint 3
SPO3
Setpoint temperature in
°
C or
°
F. Default is 40
°
C or 104
°
F.
Setpoint 4
SPO4
Setpoint temperature in
°
C or
°
F. Default is 40
°
C or 104
°
F.
Setpoint 5
SPO5
Setpoint temperature in
°
C or
°
F. Default is 40
°
C or 104
°
F.
Setpoint 6
SPO6
Setpoint temperature in
°
C or
°
F. Default is 40
°
C or 104
°
F.
Setpoint 7
SPO7
Setpoint temperature in
°
C or
°
F. Default is 40
°
C or 104
°
F.
Setpoint 8
SPO8
Setpoint temperature in
°
C or
°
F. Default is 40
°
C or 104
°
F.
Example:
PROG:DEF “My_Program”,2,0,5,1,0,3,0.0,50.0,100.0
This command is conditionally password protected.
The query returns the values of the parameters of the selected program in a
comma separated list.
PROGram:EXPlicit:DELete <name>
Delete the specified program.
<name> is the name of an existing program enclosed in quotation marks.
Since at least one program must be defined, a new default program will be
created if necessary.
This command is conditionally password protected.
PROGram:EXPlicit:EXEcute <name>
Run the specified program.
<name> is the name of an existing program enclosed in quotation marks.
No program may be running, otherwise the command will fail.