Calibrate the Product
71
Input Module Calibration
This section provides instructions to calibrate and adjust the Input Module in the
7109A-P and 6109A-P. Table 11 lists the required equipment.
Table 11. Equipment for Input Module Calibration
Equipment
Specifications
Suggested Model
0
Ω
4-wire short
0.0005
Ω
—
25
Ω
4-wire resistor
0.0005
Ω
Fluke
742A-25
100
Ω
4-wire resistor
0.001
Ω
Fluke
742A-100
200
Ω
4-wire resistor
0.002
Ω
—
400
Ω
4-wire resistor
0.004
Ω
—
Multifunction calibrator
mV: 2.0
μ
V
mA: 40 ppm + 0.4
μ
A
Fluke Calibration 5730A
Type E thermocouple
0.06
°
C at 25
°
C or 35
°
C
—
Calibration bath
0.1
°
C at 25
°
C or 35
°
C
Fluke Calibration 7109A or
6109A
Reference PRT
0.008
°
C
Fluke Calibration 5628
PRT readout
20 ppm
Fluke Calibration 1560/2560
Suitable test leads, connectors, and adapter cables are also necessary.