SYSTem:COMMunicate:SERial:LINefeed <boolean>
SYSTem:COMMunicate:SERial:LINefeed?
Enable LF termination character.
<boolean> is 0 (OFF) or 1 (ON). When this setting is on, CR and LF are both
transmitted at the end of every response. If the setting is OFF, only CR is
transmitted. The default is OFF.
SYSTem:DATE <year>,<month>,<day>
SYSTem:DATE?
Set the clock date.
This command is conditionally password protected.
SYSTem:ERRor[:NEXT]?
Return and clear the latest error from the error queue.
The response is an error code and message, separated by a comma.
Example:
-100,”Command error”
If the error queue is empty, the message is “No error”.
Example:
0,”No error”
SYSTem:KLOCk <boolean>
SYSTem:KLOCk?
Disable the front panel keys.
<boolean> is 0 (OFF, unlocked) or 1 (ON, locked). After power-on, the setting is
always 0.
SYSTem:PASSword:CDISable
Lock protected commands.
At power-on, all protected commands are locked.
SYSTem:PASSword[:CENable] <password>
Unlock protected commands.
<password> is the user password. The default password is 1234.
SYSTem:PASSword:CENable:STATe
Return the protection state.
The response is 0 (locked) or 1 (unlocked).