Portable Calibration Baths
Menu Guide
39
•
Heating is the relative size of the signal that controls the heating devices. A
negative number indicates cooling.
•
Cutout Code indicates the specific cause of a cutout event. This number is
normally 0.
•
Stir Tachometer shows the actual speed of the stir motor in RPM.
•
Stir Power is the relative size of the signal controlling the stir motor speed.
Because stir power increases with load, it can reveal a change in fluid
viscosity. An increase of 1 % or greater suggests that fluid should be checked
and perhaps replaced.
•
Typical Stir Power provides a reference to determine whether the load on the
stir motor has increased significantly. The value is set in the Bath Calibration
menu.
•
Ambient Temperature shows the air temperature inside the Product.
•
Remote Error shows the latest error message if any occurred at the time of
remote operation.
(Setup > Instrument > More > Service > System Status)
Update Firmware
Update the Product firmware.
A valid firmware update file must be stored on the memory device
plugged in to the USB host port.
(Setup > Instrument > More > Service)
Memory
Manage data stored in memory.
The Memory screen shows how many user-defined programs and test
reports are stored in memory.
(Setup > Instrument > More > Service > Memory)
Clear Data
Clear all setpoint programs and test reports from memory. The default
setpoint program remains in memory.
This function requires the password.
(Setup > Instrument > More > Service > Memory)
Clear Settings
Change all settings to factory defaults.
The settings affected are preset setpoints, operation settings, and
instrument settings. Password settings, control parameters, and
calibration parameters are not affected.
This function requires the password.
(Setup > Instrument > More > Service)
Bath Calibration
Adjust the temperature accuracy of the Product.
This menu requires the password to change any of the settings.
www.
.com
1.800.561.8187