10 - 10
91/92/96/99/105
Users Manual
Safety
Designed to Installation Category III per IEC 1010,
CSA C22.2 (including approval), and UL1244
(including approval) for 600V measurements on
industrial power distributions systems.
MAX. Input voltage on ................300V rms, 600V rms with
INPUT A or INPUT B
10:1 probe
MAX. Input voltage on EXTernal input.................300 V rms
Surge Protection ........................4 kV, 6 kV with 10:1 probe
per IEC 664 and ANSI/IEEE C62.41 test method
Maximum Isolation to Earth.....600V rms from any terminal
Optical Interface ................. Optically Isolated to 600V rms
ACCESSORY INFORMATION
Passive Probe 10:1 (PM8918)
Attenuation: x10
Useful Bandwidth: 100 MHz
Input resistance: 10 M
Ω
Input capacitance: 15 pF
Cable length: 1.5m
Max. signal Voltage: 600V rms
Max. surge Voltage: 6 kV
Max. Voltage to Ground: 600V rms
Double insulated per IEC1010-1 for 600V.
Optically Isolated RS-232 Adapter /Cable (PM9080/001)
Serial Printer Interface
............. EPSON FX/LQ,
HP Thinkjet, HP Laserjet
Print screens and waveforms or
log meter and cursor readings
Full
PC
Interface
..............Transfer
setups,
waveforms, measurements, etc.
Capture Screen ...........With optional FlukeView™ software
91
92
91
92
96
Содержание 105 Series II
Страница 6: ...V ...
Страница 14: ...XIII Figure 2 Demo Board Non conductive surface ...
Страница 15: ...XIV 91 92 96 99 105 Users Manual ...
Страница 17: ...XVI 91 92 96 99 105 Users Manual ...
Страница 204: ...10 34 91 92 96 99 105 Users Manual Figure 10G 5 EXT mV menu MORE EXT mV ALERT ZERO FAST SMOOTH RESET READING ...
Страница 205: ...Appendixes 10 35 Figure 10G 6a MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6b ...
Страница 206: ...10 36 91 92 96 99 105 Users Manual Figure 10G 6b MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6a ...
Страница 212: ...10 42 91 92 96 99 105 Users Manual Figure 10G 13 SPECIAL FUNCTION menu MENU TEST MENU GENERATE MATH COMP 92 91 96 ...