Using the Dual Display Mode Functions
4 - 13
Readings as a Percent Change from Reference Point
(Zero %
∆
)
Select "ZERO %
∆
" to set the present upper reading as the
relative reference. Subsequent readings are displayed as a
percentage of variation (delta) from this point.
The figure below shows percentage change (tolerance)
when measuring a resistor:
Readings Displayed as a Percent of Scale (0%-100%)
Scaling starts automatically when both 0% and 100%
values have been entered as the scaled reference values.
Use the following procedure (example for Meter mode):
1. From the SCALE READING list box, highlight 0%-100%
and press
.
2. Use
to highlight REFERENCE 100% and
press .
3. Measure the value that will represent 100%.
4. Highlight NOW (as the present 100% value) and press
to save the displayed value as 100%.
5. Repeat actions 2 to 4 for the 0% value.
6. Press
(CLOSE) to activate scaling and remove
the window.
The present measurement result with respect to the
defined range with a 0% and 100% value appears now as a
scaled % readout.
Содержание 105 Series II
Страница 6: ...V ...
Страница 14: ...XIII Figure 2 Demo Board Non conductive surface ...
Страница 15: ...XIV 91 92 96 99 105 Users Manual ...
Страница 17: ...XVI 91 92 96 99 105 Users Manual ...
Страница 204: ...10 34 91 92 96 99 105 Users Manual Figure 10G 5 EXT mV menu MORE EXT mV ALERT ZERO FAST SMOOTH RESET READING ...
Страница 205: ...Appendixes 10 35 Figure 10G 6a MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6b ...
Страница 206: ...10 36 91 92 96 99 105 Users Manual Figure 10G 6b MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6a ...
Страница 212: ...10 42 91 92 96 99 105 Users Manual Figure 10G 13 SPECIAL FUNCTION menu MENU TEST MENU GENERATE MATH COMP 92 91 96 ...