Appendixes
10 - 7
Ω
MODE
(external 4-mm banana jack inputs)
Ranges ...... 30
Ω
, 300
Ω
, 3 k
Ω
, 30 k
Ω
, 300 k
Ω
, 3 M
Ω
, 30 M
Ω
Accuracy (300
Ω
to 30 M
Ω
)
....................... ±
(0.5% +5 counts)
Accuracy (30
Ω
)
......................................... ±
(2.5% +25 counts)
Full Scale Reading........................................... 3000 counts
Measurement Current ........... 500
µ
A, 500
µ
A, 70
µ
A, 7
µ
A,
700 nA, 70 nA, 70 nA
Open Circuit Voltage ..................................................... <4V
Full Scale Voltage ............<250 mV to 3 M
Ω
, <2V to 30 M
Ω
Continuity Beeper ................................ audible for readings
<5% of selected range
DIODE TEST MODE
(external 4 mm banana jack inputs)
Open Circuit Voltage
............................................................≤
4V
Full Scale Voltage .................................................... 2.800V
Measurement Current .............................................. 0.5 mA
Continuity Beeper ......................... audible for readings <1V
EXTERNAL mV MODE
Displays up to four readings and a full size waveform of the
4 mm banana jack input.
DC Voltage
Ranges ..............................................................300 mV, 3V
Accuracy
...................................................... ±
(0.5% +5 counts)
Full Scale Reading........................................... 3000 counts
Normal Mode Rejection ...................>50 dB at 50 or 60 Hz
Common Mode Rejection .............. >100 dB at DC, 50, 60,
or 400 Hz
AC or AC+DC True RMS Voltage
Ranges ............................................................. 300 mV, 3V
Accuracy .......................................... DC, 50 Hz, and 60 Hz.
(valid from >5% of range)
±
(2% +15 counts)
OL at >2500 counts in 3V range
Full Scale Reading........................................... 3000 counts
Crest Factor .....Automatic ranging on crest factor overload
Useful Bandwidth .........................................................5kHz
Common Mode Rejection Ratio ..........>60 dB, DC to 60 Hz
Additional EXTernal mV mode measurements
Frequency, RPM, Duty Cycle and Pulse Width
measurements and waveforms can also be made with
reduced accuracy.
°C / °F TEMPERATURE SCALING .................. for optional
temperature probes
Scaling ............................................................ 1 mV/degree
AMPERE SCALING .................. for optional current probes
Scaling ...... 1 mV/A, 10 mV/A, 100 mV/A, 1 V/A (1 mV/mA)
for DC, AC rms, or AC+DC rms
Содержание 105 Series II
Страница 6: ...V ...
Страница 14: ...XIII Figure 2 Demo Board Non conductive surface ...
Страница 15: ...XIV 91 92 96 99 105 Users Manual ...
Страница 17: ...XVI 91 92 96 99 105 Users Manual ...
Страница 204: ...10 34 91 92 96 99 105 Users Manual Figure 10G 5 EXT mV menu MORE EXT mV ALERT ZERO FAST SMOOTH RESET READING ...
Страница 205: ...Appendixes 10 35 Figure 10G 6a MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6b ...
Страница 206: ...10 36 91 92 96 99 105 Users Manual Figure 10G 6b MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6a ...
Страница 212: ...10 42 91 92 96 99 105 Users Manual Figure 10G 13 SPECIAL FUNCTION menu MENU TEST MENU GENERATE MATH COMP 92 91 96 ...