Using the Dual Display Mode Functions
4 - 5
The measurement result is displayed as a 3 2/3-digit
display. The ranges are 30
Ω
(manual only), 300
Ω
, 3 k
Ω
,
30 k
Ω
, 300 k
Ω
, 3 M
Ω
, and 30 M
Ω
. A beeper alert is
generated at <5% of a selected range in Manual Range
mode. The beeper alert is turned on with ALERT ))).
Diode test mode. A diode connected to the red
and black 4-mm banana jack inputs is tested. The
measurement current through the diode is 0.5 mA. The red
banana jack is the high input, and the black banana jack
(COM) is the low.
The result is displayed as the forward or reverse voltage of
the diode. A beeper alert is generated at forward voltage.
The beeper alert is turned on with ALERT ))).
The time base is fixed at 5 ms/DIV and the attenuator at 3V.
NOTE
In the diode and resistance modes, the black COMmon
banana jack is not connected internally to the common of
BNC INPUTS A and B.
Содержание 105 Series II
Страница 6: ...V ...
Страница 14: ...XIII Figure 2 Demo Board Non conductive surface ...
Страница 15: ...XIV 91 92 96 99 105 Users Manual ...
Страница 17: ...XVI 91 92 96 99 105 Users Manual ...
Страница 204: ...10 34 91 92 96 99 105 Users Manual Figure 10G 5 EXT mV menu MORE EXT mV ALERT ZERO FAST SMOOTH RESET READING ...
Страница 205: ...Appendixes 10 35 Figure 10G 6a MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6b ...
Страница 206: ...10 36 91 92 96 99 105 Users Manual Figure 10G 6b MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6a ...