background image

 

November 2012 

© 2012 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. 
All product names are trademarks of their respective companies. 

CNX v3000 

Wireless AC Voltage 

 
 
 

Calibration Manual 

 

Summary of Contents for CNX v3000

Page 1: ...012 2012 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies CNX v3000 Wireless AC Voltage Calibrat...

Page 2: ...r nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB...

Page 3: ...equired Equipment 5 Performance Tests 5 Test the Display 5 Backlight 6 Keypad Test 6 AC Voltage Test 6 Before Calibration Adjustment 7 Maintenance Mode 7 Password Entry 8 Change the Password 8 Restore...

Page 4: ...CNX v3000 Calibration Manual ii...

Page 5: ...iii List of Tables Table Title Page 1 Symbols 3 2 Required Equipment 5 3 Performance Tests 6 4 Calibration Adjustment 10 5 User Replaceable Parts 12...

Page 6: ...CNX v3000 Calibration Manual iv...

Page 7: ...v List of Figures Figure Title Page 1 All Segments of the Display 5 2 Performance Test Connections 7 3 Calibration Password Reset 9 4 Battery Replacement 12...

Page 8: ...CNX v3000 Calibration Manual vi...

Page 9: ...luke call one of the following telephone numbers Technical Support USA 1 800 44 FLUKE 1 800 443 5853 Calibration Repair USA 1 888 99 FLUKE 1 888 993 5853 Canada 1 800 36 FLUKE 1 800 363 5853 Europe 31...

Page 10: ...urements Have an approved technician repair the Product Use only specified replacement parts Do not exceed the Measurement Category CAT rating of the lowest rated individual component of a Product pro...

Page 11: ...o test and measuring circuits connected directly to utilization points socket outlets and similar points of the low voltage MAINS installation CAT III Measurement Category III is applicable to test an...

Page 12: ...4 GHz ISM Band RF Communication Range 20 m 65 61 ft Operating Temperature 10 C to 50 C 14 F to 122 F Storage Temperature 40 C to 60 C 40 F to 140 F with batteries removed Operating Humidity 90 from 0...

Page 13: ...test procedures unless the Product is fully assembled The performance tests verify the full operation of the Product and measure the accuracy of each function against Product specifications If the Pro...

Page 14: ...up the Calibrator as necessary Refer to its specifications 4 Let the temperature of the unit under test UUT become stable to room temperature To do the ac voltage test 1 Connect the Product to Calibra...

Page 15: ...Adjustment Before the Product calibration can be adjusted you must go through the maintenance mode menu and enter your password Maintenance Mode The Product maintenance mode can be used to set differ...

Page 16: ...correct password is entered C 01 is shown If the incorrect password is entered is shown and the password must be correctly entered Change the Password Note If you change the password and then lose it...

Page 17: ...r handling them 3 Using the procedures packaging and bench techniques that are recommended The following practices should be followed to minimize damage to S S static sensitive devices 1 MINIMIZE HAND...

Page 18: ...E EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC FREE WORK STATION PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S S DEVICES 9 HANDLE S S DEVICES ONLY AT A STATIC FREE...

Page 19: ...Phillips screwdriver remove the bottom case screws Two of the screws are inside of the battery door 3 Keep the pca in the top case 4 Apply 3 0 V across the battery contacts on the pca Note the polarit...

Page 20: ...on After confirmation the Product goes to the subsequent calibration step Note After you push wait until the calibration step number advances before you change the Calibrator source Some adjustment st...

Page 21: ...uct or to equipment under test Remove batteries to prevent battery leakage and damage to the Product if it is not used for an extended period Be sure that the battery polarity is correct to prevent ba...

Page 22: ...Part Number Description Qty 4130305 FLK 3000 2003 DOOR BATTERY 1 1881997 TPAK Meter Hanging Kit 1 2061552 TL224 4201 175 263 011 TEST LEADS RA2S 2 1610159 AC285 5001 01 175 276 011 AC285 LARGE ALLIGA...