10 - 6
91/92/96/99/105
Users Manual
METER MODE
Displays up to four readings and a full screen scope
waveform of INPUT A.
DC Voltage
Ranges
direct input .................. 100 mV, 300 mV, 1V, 3V, 10V, 30V,
100V, 300V
with 10:1 probe .......... 1V, 3V, 10V, 30V, 100V, 300V, 1 kV,
3 kV (limit 600V)
Accuracy
......................................................±
(0.5% +5 counts)
Full Scale Reading ............................. 1000 or 3000 counts
Normal Mode Rejection ................... >50 dB at 50 or 60 Hz
Common Mode Rejection ...............>100 dB at DC, 50, 60,
or 400 Hz
AC or AC+DC True RMS Voltage
Ranges
direct input ................. 100 mV, 300 mV, 1V, 3V, 10V, 30V,
100V, 250V
with 10:1 probe .......... 1V, 3V, 10V, 30V, 100V, 300V, 1 kV,
2.5 kV (limit 600V)
Accuracy .................... 50 Hz to 60 Hz
±
(1% +10 counts)
(valid from
1 Hz [20 Hz] to 20 kHz
±
(2% +15 counts)
5% range)
1 Hz [5 Hz] to 1 MHz
±
(3% +20 counts)
1 Hz [5 Hz] to 5 MHz
±
(10% +25 counts)
•
[values] for "Vrms AC (no DC)", AC coupled input.
•
In "AC+DC", the DC component is included in the
measurement result and does not influence the accuracy.
Full Scale Reading ............................. 1000 or 3000 counts
Crest Factor ..... Automatic ranging on crest factor overload
Common Mode Rejection Ratio.......... >60 dB, DC to 60 Hz
Additional Meter Mode Measurements
FREQUENCY .............................................. 1 Hz to 5 MHz
Resolution.................................................................4 digits
Accuracy
......................................................±
(0.5% +2 counts)
RPM ..................... selectable: 1 pulse/rev. or 1 pulse/2 rev.
Range .......... 60RPM (120 @ 1 pulse/2 rev.) to 99.99kRPM
Resolution......................1 RPM (10RPM for RPM >10,000)
Accuracy
.......................................................±
(1% +10 counts)
DUTY CYCLE .............................positive or negative pulse
Range ...........................................................2.0% to 98.0%
Resolution.................................................................... 0.1%
Accuracy (logic or pulse waveforms)
.......±
(0.5% +2 counts)
PULSE WIDTH ..........................positive or negative pulse.
Range ........................................................ 250
µ
s to 50 ms
Resolution.................................................................3 digits
Accuracy
......................................................±
(0.5% +2 counts)
dBV .............................................................1 volt reference
dBm ........................ reference @ 50, 60, 75, 93, 110, 125,
135, 150, 250, 300, 500, 600, 800,
900, 1000 or 1200
Ω
dBW or WATTS .................. 1, 2, 4, 8, 16 or 50
Ω
reference
Содержание 105 Series II
Страница 6: ...V ...
Страница 14: ...XIII Figure 2 Demo Board Non conductive surface ...
Страница 15: ...XIV 91 92 96 99 105 Users Manual ...
Страница 17: ...XVI 91 92 96 99 105 Users Manual ...
Страница 204: ...10 34 91 92 96 99 105 Users Manual Figure 10G 5 EXT mV menu MORE EXT mV ALERT ZERO FAST SMOOTH RESET READING ...
Страница 205: ...Appendixes 10 35 Figure 10G 6a MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6b ...
Страница 206: ...10 36 91 92 96 99 105 Users Manual Figure 10G 6b MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6a ...
Страница 212: ...10 42 91 92 96 99 105 Users Manual Figure 10G 13 SPECIAL FUNCTION menu MENU TEST MENU GENERATE MATH COMP 92 91 96 ...