7 - 8
91/92/96/99/105
Users Manual
➀
Select Continuous Auto Set.
➁
➂
Open the Math menu.
➃
➪
Select each item.
Store the resulting waveform in Destination memory 1.
➄
Close the window.
In Figure 7-7, INPUT A represents the current measure-
ment, INPUT B the voltage measurement, and Destination
memory 1 the product (equals power).
In this example the following scales are used:
INPUT A:
20 A/DIV
INPUT B:
200 V/DIV
Memory 1:
4 kW/DIV
Figure 7-7 Power measurement
Shift the resulting waveform over the display as follows:
➅
➆
:
move
1
➪
➇
Move the resulting waveform.
Содержание 105 Series II