5 - 4
91/92/96/99/105
Users Manual
Capturing Glitches
Press
on INPUT A to turn GLITCH DETECTion on.
This function displays the events that occur between two
samples on INPUT A. INPUT B is turned off. The events
can be glitches or other asynchronous waveforms that are
40 ns (nanoseconds) or wider.
NOTE
GLITCH detection and AVERAGE are mutually exclusive.
Reversing the Polarity of the Displayed Waveform
Press
to open the MORE INPUT window. Highlight
INPUT A (or INPUT B) and press
to open the list box.
You can turn on and off an input and invert the INPUT A
and INPUT B display.
Invert reverses the polarity of the displayed waveform. This
is identified by the trace identifier
on the display. For
example, a negative-going waveform would be displayed
as positive-going, providing a more meaningful viewing
perspective in some cases.
ADJUSTING THE AMPLITUDE
You can use
for INPUT A or INPUT B
to
adjust input waveform attenuation. Sixteen ranges are
available. A 1:1 probe can accommodate 1 mV to a
maximum of 100V per division, and a 10:1 probe can
accommodate 10 mV to a maximum of 1 kV per division.
The amplitude can be changed in increments of 1, 2, and 5
(e.g., 100 mV, 200 mV, 500 mV). The upper row of the top
display shows the range and the type of probe for each
input.
When a 10:1 probe is used with the 10 mV or 20 mV range,
averaging is turned on automatically. (AVG 4 appears on
the display.)
91
Содержание 105 Series II
Страница 6: ...V ...
Страница 14: ...XIII Figure 2 Demo Board Non conductive surface ...
Страница 15: ...XIV 91 92 96 99 105 Users Manual ...
Страница 17: ...XVI 91 92 96 99 105 Users Manual ...
Страница 204: ...10 34 91 92 96 99 105 Users Manual Figure 10G 5 EXT mV menu MORE EXT mV ALERT ZERO FAST SMOOTH RESET READING ...
Страница 205: ...Appendixes 10 35 Figure 10G 6a MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6b ...
Страница 206: ...10 36 91 92 96 99 105 Users Manual Figure 10G 6b MEASURE menu MORE MEASURE V dB TIME Hz See Figure 10G 6a ...
Страница 212: ...10 42 91 92 96 99 105 Users Manual Figure 10G 13 SPECIAL FUNCTION menu MENU TEST MENU GENERATE MATH COMP 92 91 96 ...