OPERATING INSTRUCTIONS

MODEL 2705B

DIGITAL MULTIMETER

10A

COM

A mA

V

2705B

SAFETY INFORMATION

The following safety information must be observed to ensure maximum personal
safety during the operation at this meter:

Use the meter only as specified in this manual or the protection provided by the
meter might be impaired.

Test the meter on a known voltage before using it to determine if hazardous
voltage is present.

Do not use the meter if the meter or test leads look damaged, or if you suspect
that the meter is not operating properly.

Never ground yourself when taking electrical measurements. Do not touch
exposed metal pipes, outlets, fixtures, etc., which might be at ground potential.
Keep your body isolated from ground by using dry clothing, rubber shoes, rubber
mats, or any approved insulating material.

Turn off power to the circuit under test before cutting, unsoldering, or breaking
the circuit. Small amounts of current can be dangerous.

Use caution when working above 60V dc or 30V ac rms. Such voltages pose a
shock hazard.

When using the probes, keep your fingers behind the finger guards on the
probes.

Measuring voltage which exceeds the limits of the multimeter may damage the
meter and expose the operator to a shock hazard. Always recognize the meter
voltage limits as stated on the front of the meter.

SPECIFICATIONS

Display:

3¾ digit liquid crystal display (LCD) with a maximum reading of 3999.

Polarity:

Automatic, positive implied, negative polarity indication.

Overrange:

(OL) or (-OL) is displayed.

Zero:

Automatic.

Low battery indication

: The

"

"

is displayed when the battery voltage drops

below the operating level.

Measurement rate:

2 times per second, nominal.

Auto power off:

Approx. 10 minutes.

Operating environment:

0

to 50

at < 70% relative humidity.

Storage temperature:

-20

to 60

, 0 to 80% relative humidity.

Accuracy:

Stated accuracy at 23

±5

, < 75% relative humidity.

Temperature Coefficient: 0.1 x (specified accuracy) per

. (

to 18

, 28

to 50

).

Altitude:

6561.7 feet (2000m).

Power:

Single standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22.

Battery life:

200 hours typical with carbon-zinc.

Dimensions:

165mm (H) x78mm (W) x42.5mm (D).

Weight:

Approx. 10.0 oz.(285g) including holster.

Accessories:

One set test leads, one spare fuse, 9V battery (installed), and

DC VOLTS

Ranges:

400mV, 4V, 40V, 400V, 1000V

Resolution:

0.1mV

Accuracy:

± ( 1 . 0 % r d g + 2 d g t s )

Input impedance:

400mV: >100M

Ω

; 4V:10M

Ω

; 40V ~ 1000V:9.1M

Ω

Overload protection:

1000VDC or 750VAC rms

AC VOLTS

(50Hz - 500Hz)

Ranges:

400mV, 4V, 40V, 400V, 750V

Resolution:

0.1mV

Accuracy:

± ( 2.0% rdg + 5 dgts) 50 ~100Hz on 400mV range
± ( 2.0% rdg + 5 dgts)

Input impedance:

400mV: >100M

Ω

; 4V:10M

Ω

; 40V ~750V:9.1M

Ω

Overload protection:

1000VDC or 750VAC rms

CURRENT

Ranges:

400uA, 4000uA, 40mA, 400mA, 10A

Resolution:

0.1uA

DC accuracy:

± ( 2.0% rdg + 2 dgts) on 400uA to 400mA ranges
± ( 3.0% rdg + 3 dgts) on 10A range

AC accuracy:

(50Hz ~ 500Hz)

± ( 2.5% rdg + 5 dgts) on 400uA to 400mA ranges
± ( 3.5% rdg + 5 dgts) on 10A range

Voltage burden:

0.2V on 400uA, 40mA, 10A ranges

2V on 4000uA, 400mA ranges

Input protection:

0.5A/500V fast blow ceramic fuse

10A/600V fast blow ceramic fuse

10A Input

: 10A for 60 seconds maximum followed by a 10 minute cooling period

RESISTANCE

Ranges:

400

Ω

, 4k

Ω

, 40k

Ω

, 400k

Ω

, 4M

Ω

, 40M

Ω

Resolution:

0.1

Ω

Accuracy:

± ( 1.5% rdg + 4 dgts) on 400

Ω

to 400k

Ω

ranges

± ( 2.5% rdg +4 dgt) on 4M

Ω

range

± ( 5.0% rdg + 5 dgt) on 40M

Ω

range

Open circuit volts:

-0.45Vdc (-1.2Vdc on 400

Ω

range)

Overload protection:

500VDC or AC rms

DIODE TEST

Test current:

1.0mA (approximate)

Accuracy:

±(3.0% rdg + 3 dgts)

Resolution:

10mV

Audible indication:

<0.25V

Open circuit volts:

3.0Vdc typical

Overload protection:

500VDC or AC rms

CONTINUITY

Audible indication:

Less than 25

Ω

Response time:

500ms

Overload protection:

500VDC or AC rms

OPERATION

Before taking any measurements, read the Safety Information Section. Always
examine the instrument for damage, contamination (excessive dirt, grease, etc.)
and defects. Examine the test leads for cracked or frayed insulation. If any
abnormal conditions exist do not attempt to make any measurements.

the user when the test lead is in the current

make a voltage measurement. Another safety
u.

ading on display, and release it by pressing

e mode. In manual range, you select and lock

lect a range:

selected range. Subsequently pressing the

nge in sequence from the lowest to highest

s to return to the Autorange Mode.

jack and the black test lead to the ”COM”

the desired voltage type (AC or DC) and

not known, set switch to the highest range

ading is obtained.

ce or circuit being measured.

negative polarity; positive polarity is implied.

A, mA or 10A) jack and the black test lead to

the DC or AC ranges.
der test and open the normal circuit path

taken. Connect the meter in series with the

om the display.

the desired resistance range.

t under test.

V

Ω

" jack and the black test lead to the "COM"

ts of measurements and read the value from

V

Ω

” jack and the black test lead to

the “

“ position.

test. External voltage across the

gs.

ard-voltage drop is about 0.6V (typical for a

od, “OL” is displayed. If the diode is shorted,

yed.

yed in both directions.

V.

position.

r test. External Voltage across the

g.

points at which continuity is to be tested. The

is less than approximately 25

Ω

.

es.

tton to restart the meter, and the reading of

in the display.

er Off Feature:

while rotating function switch from off to

e auto power off feature is disabled.

from the LCD

.

MAINTENANCE

WARNING

Remove test leads before changing battery or fuse or
performing any servicing.

Battery Replacement

Power is supplied by a 9 volt battery. (NEDA 1604, IEC 6F22). The

"

"

appears

on the LCD display when replacement is needed. To replace the battery, remove
the three screws from the back of the meter and lift off the front case. Remove
the battery from case bottom.

Fuse Replacement

If no current measurements are possible. Check for a blown overload protection
fuse. For access to fuses, remove the three screws from the back of the meter
and lift off the front case. Replace F1 only with the original type 0.5A/500V, fast
acting ceramic fuse, 6.35x32mm Replace F2 only with the original type
10A/600V, fast acting ceramic fuse, 6.35x25.4mm.

Cleaning

Wipe the case with a damp cloth and mild detergent. Do not use abrasives or
solvents. Dirt or moisture in the terminals can affect readings.

Safety:

Conforms to IEC61010-1 (EN61010-1), CATII 1000V, CATIII 600V, Class

II, Pollution degree 2 Indoor use.

CATII:

Is for measurements performed on circuits directly connected to the

low-voltage installation.

CAT III:

Is for measurements performed in the building installation.

EMC:

Conforms to EN61326.

The symbols used on this instrument are: