Troubleshooting
58
Error
Symptom
Possible causes and corrective measures
E-33
Rotor stops without brake
to standstill or does not
start.
Overpressure in the refrigeration system
1.
Turn instrument off and pull main plug.
1. Check and clean ventilation slots if necessary.
2. After about 60 min you can restart the instrument.
Observe the max. permissible environmental temperature!
If the error persists, contact service.
E-34
Rotor stops without brake
to standstill or does not
start.
Overvoltage in the intermediate circuit.
Switch the instrument off and on again.
If the error persists, contact service.
E-35
Rotor stops without brake
to standstill or does not
start.
Overcurrent in the intermediate circuit.
Switch the instrument off and on again. If the error persists,
contact service.
E-36
|
E-38
Rotor stops without brake
to standstill.
Instrument cannot be
operated.
Overcurrent or error in current measurement.
Switch the instrument off and on again.
If the error persists, contact service.
E-39
Rotor stops without brake
to standstill.
Instrument cannot be
operated.
Speed control measurement exceeds permissible rotor
speed.
Switch the instrument off and on again.
If the error persists, contact service.
Содержание accuSpin 3
Страница 1: ...accuSpin 3 3R OPERATING INSTRUCTIONS...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the accuSpin 12 Notes...
Страница 20: ...Rotors and accessories 16...
Страница 32: ...Rotors and accessories 28 Notes...
Страница 38: ...Operation 34...
Страница 68: ...64 Notes...
Страница 70: ...66 Acceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395 time s speed rpm 9 1...
Страница 71: ...67 speed rpm time s 1 9 Deceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395...
Страница 72: ...68 Acceleration profiles Micro liter Rotor 7500 3337 speed rpm time s...
Страница 73: ...69 Deceleration profiles Micro liter Rotor 7500 3337 speed rpm time s 9 1...
Страница 76: ...72...