Troubleshooting
55
Error
Symptom
Possible causes and corrective measures
E-19
Instrument stops with
deceleration to standstill
after short starting-up.
No rotor present or rotor identification impossible.
A) Check whether a certified rotor is inserted.
B) Please take care of the readability of the inscription of
the swinging bucket rotor cross installed.
(rotor identification must show to the chamber bottom)
C) For the swinging bucket rotor the set bucket type must be
permitted for the operation in the respective rotor.
Please note the hints of chapter "Bucket selection of
swinging bucket rotors" (page 35) and compare the
permitted rotors and rotor buckets in "table 1" on page
16.
D) Following a brief power failure, the rotor could not be
identified. Switch the instrument off and on again using
the mains switch.
E-20
Instrument does not start or
brakes to standstill.
Rotor cannot be identified.
A) Check whether a certified rotor is inserted.
B) Check whether the rotor is completely fastened onto the
motor shaft. For this release the rotor fastening until you
can lift and depress it easily. Tighten the rotor fastening
once again now.
Содержание accuSpin 3
Страница 1: ...accuSpin 3 3R OPERATING INSTRUCTIONS...
Страница 10: ...For your safety 6 Notes...
Страница 16: ...Description of the accuSpin 12 Notes...
Страница 20: ...Rotors and accessories 16...
Страница 32: ...Rotors and accessories 28 Notes...
Страница 38: ...Operation 34...
Страница 68: ...64 Notes...
Страница 70: ...66 Acceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395 time s speed rpm 9 1...
Страница 71: ...67 speed rpm time s 1 9 Deceleration profiles Swinging bucket rotor with rectangular buckets 7500 4395...
Страница 72: ...68 Acceleration profiles Micro liter Rotor 7500 3337 speed rpm time s...
Страница 73: ...69 Deceleration profiles Micro liter Rotor 7500 3337 speed rpm time s 9 1...
Страница 76: ...72...